41 PCI-1706U User Manual
Appendix C Calibration
Step 2: Calibrate gain.
Item 1: Select a gain in the combo box and connect a specific signal to the AI chan-
nel(s).
Item 2: Click "Start" button to start calibrating.
Item 3: Display some values during calibration process. These values are for internal
use, send them to Advantech if calibration fails.
Item 4: Display the calibration result. If success the status is Done otherwise is
“Failed”.
Item 5: Select another gain to calibrate and repeat the 1-4 above.
Note! The signal voltage is different at different gain, see the red box marked
"1" for the correct value.